• 黄金城集团

    DNA Extraction from Modern Dental Plaque on Gauze

    Franziska Aron, Irina Velsko

    Published: 2022-04-26 DOI: 10.17504/protocols.io.eq2lypejrlx9/v1
    AI 解读
    浏览 1442

    Abstract

    Protocol for DNA extraction from modern dental plaque dried on gauze samples for Illumina sequencing. This protocol uses the Qiagen PowerSoil DNA extraction kit, but applies modifications to improve DNA recovery. Specifically, solution C2 is not used to avoid uneccessary DNA loss during inhibitor removal.

    Before start

    Planning

    This protocol requires the use of a Biological Safety Laboratory level S2 due to handling of human substrates.

    This protocol takes ~1 day.

    Prepare a cool tube-rack for 1.5ml and 2ml tubes by placing it at +4°C for the DNA-clean up. This will speed up reaction time with a 'cold start'.

    Modern dental plaque samples should be stored at least at +4°C or preferably at -20°C to preserved DNA after sampling.

    Check waste disposal guidance for all reagents in this protocol against your corresponding laboratory regulations.

    Equipment

    Make sure all necessary equipment is available (see Materials).

    Abbreviations

    HPLC = High Performance Liquid Chromatography (i.e., HPLC-grade water)

    Controls

    Consider taking along a positive control (sample of known performance) and a negative control (tube with HLPC-grade water instead of DNA) in order to assess the performance of the protocol and the level of background contamination. Take into account these two extra samples in your calculations for buffer preparations.

    Steps

    Preparation Day 1

    1.

    Place a 1.5ml/2ml tube rack at 4°C for the incubation steps on the final days

    2.

    Label for each sample one 2ml-Eppendorf Safe-Lock DNA LoBind-Tube

    and one 1.5ml-Eppendorf Safe-Lock DNA LoBind-Tube

    3.

    Place solution C6 in an incubator to pre-heat it to 37°C.

    Preparation for Final Day

    4.

    Retrieve the spin filter columns (provided in the kit) from fridge to warm them up to -20Room temperature for use at step 25

    Note
    "provided in the kit" refers to: DNeasy PowerSoil Kit (100) Cat No./ID: 12888-100A spin filter column is a spin filter unit sitting in a collection column. The spin filter unit contains the membrane.

    5.

    Label for each sample one Power Bead tube, one spin filter column and three 2ml collection tubes (all provided in the kit).

    Note
    Be sure to label both the top and sides of the PowerSoil bead tube. The labeling on the top may be rubbed off during bead-beating with the Omni BeadRuptor.

    Sample Preparation

    6.

    Sterilize scissors by wiping with 75% ethanol, and then wiping them with HPLC water, then wipe dry. Use the scissors to clip a small piece of gauze with plaque.

    Note
    Do not take more than half of the plaque, in case more is needed for another extraction later.

    7.

    Place the piece of gauze with plaque in a PowerSoil bead tube using forceps sterilized by wiping with 75% ethanol, and then wiping them with sterile water.

    Safety information
    Change gloves and sterilize scissors and forceps between each sample.

    8.

    Make two extraction blank PowerSoil bead tubes per extraction batch, one with nothing added, and a second with a small section of sterile gauze added.

    DNA_Clean_up

    9.

    Invert the PowerSoil bead tubes to mix and fully submerge gauze with plaque.

    10.

    Add 60µL C1 solution, invert several times to mix.

    11.

    Add 25µLproteinase K (10mg/mL), and rotate tubes several times to mix.

    12.

    Place tubes in an OmniBeadRuptor, secure them, and close the lid. Run the BeadRuptor at 2.9 m/s for 0h 10m 0s

    Note
    Make sure the Power Bead tubes are secure, but do not close the securing lid too tightly, 2.9 m/s is the maximum, otherwise the tubes may break.

    13.

    Centrifuge at 9400x g,0h 0m 0s for 0h 0m 30s at Room temperature

    14.

    Transfer 600µL supernatant to the 2ml collection tube (provided in the kit), but leave the gauze in the bead tube.

    Be careful not to transfer any beads or dust from the beads.

    14.1.

    Save power bead tubes at -20°C as backup.

    Note
    We save this as backup until confirmation that the extraction worked. If extraction was successful, this can be discarded.

    15.

    Add 200µL C3 solution and vortex briefly

    16.

    Incubate the tubes at 4°C for 0h 5m 0s on the pre-cooled rack (4°C )

    17.

    Centrifuge the tubes at 9400x g,0h 0m 0s for 0h 1m 0s at 4Room temperature

    18.

    Transfer 750µL of the supernatant to a clean 2ml collection tube (provided in the kit)

    18.1.

    Store remaining supernatant at -20°C

    Note
    We save this as backup until confirmation that the extraction worked. If extraction was successful, this can be discarded.

    19.

    Gently shake C4 Solution to mix

    20.

    Add 1.2mL C4 Solution to supernatant and vortex for 0h 0m 5s

    Note
    The solution should not exceed the rim of the tube.

    21.

    Bind the DNA on the membrane of spin filter unit:

    21.1.

    Load approximately 675µL into spin filter unit

    21.2.

    Centrifuge 0h 1m 0s at 9400x g,0h 0m 0s at 4Room temperature

    21.3.

    Discard flow through

    22.

    Repeat until the entire solution has been passed through the spin filter unit (2-3 times)

    23.

    Load 500µL C5 Solution into spin filter unit

    24.

    Centrifuge 0h 1m 0s 9400x g,0h 0m 0s at -20Room temperature

    25.

    Discard flow through

    26.

    Dry spin 0h 1m 0s 9400x g,0h 0m 0s at-20Room temperature

    Elution

    27.

    Place the spin filter unit into a clean 2mL collection tube (provided in the kit), or if preferred, a 1.5 mL LoBind tube

    28.

    Pipette 100µL C6 Solution (pre-warmed to 37°C) into center of the membrane in the spin filter unit and incubate for 0h 1m 0s

    29.

    Centrifuge 9400x g,0h 0m 0s for 0h 0m 30s at -20Room temperature

    30.

    Remove spin filter unit and quantify the DNA (3µL) with the Qubit ds BR Kit following the manufacturer's protocol

    30.1.

    Store the eluate at-20°C

    推荐阅读