Fluorescent Western Protocol
Lynn Doran, Steven J Burgess
Abstract
Analysis of proteins using fluorescent immunoblot.
Note:
-
The choice of secondary antibody depends on the choice of primary antibody, whether it is derived from a mouse (monoclonal) or a rabbit (polyclonal).
-
It is advisable to stick to the 800CW wavelength to avoid problems with chlorophyll autofluorescence encountered with the 680CW antibodies.
Literature:
Licor's "Fluorescent Western Blot Detection"
Licor's "Good Westerns Gone Bad"
Before start
Isolate total protein via Leaf Protein Extraction for Immunoblot (Soybean, Cowpea, Tobacco).
Quantify protein via Protein Concentration Determination using Qubit 4 Fluorometer.
Separate protein components via SDS-PAGE gel electrophoresis.
Transfer protein to a membrane via Protein Transfer using Bio-rad TransBlot Turbo.
Steps
Keep membranes in the black Western Blot incubation box for all steps, this is important after adding the secondary antibody because the signal is light-sensitive and will become bleached if not exposed to light for a long enough period.
Wet with 1x PBS for 0h 2m 0s min
Rinse membrane with dH2O
Discard PBS and incubate with 1h 0m 0s at Room temperature
Prepare 1L PBS-T solution by diluting 100 mL of 10X PBS
Incubate with primary antibody (appropriate dilution in 4°C (in cold room) with gentle agitation on a platform shaker 50rpm .
Pour off the primary antibody and rinse the membrane with PBS-T.
Cover the membrane with PBS-T, shake vigorously on a platform shaker at 50rpm. Repeat 3 times.
Create a working dilution of secondary antibody using
Incubate for 1h 0m 0s Room temperature with gentle agitation on a platform shaker.
Pour off the secondary antibody and rinse membrane with distilled water to remove residual blocking agent.
Cover the membrane with PBS-T, agitate 80rpm
Discard PBS-T. Repeat step 12 three times.
Rinse then cover the membrane with 1x PBS, 80rpm.
Proceed to imaging blot on LI-COR Odyssey CLx imaging system
Equipment
| Value | Label |
|---|---|
| Odyssey CLx | NAME |
| Imaging System | TYPE |
| LI-COR | BRAND |
| Odyssey CLx | SKU |