    LRRK2 thermal shift assay

    Deep Chatterjee, Sebastian Mathea, Stefan Knapp, Verena Dederer

    Published: 2023-10-26 DOI: 10.17504/protocols.io.kxygx3y6kg8j/v1
    AI 解读
    浏览 1319

    Abstract

    Thermal shift assay or differential scanning fluorimetry analyzes the effect of small molecules on the thermostability of a protein by gradual heat denaturation and monitoring absorption of the fluorescent dye SYPR Orange at 488 nm.

    Steps

    Fluorescent-based thermal shift assay

    1.

    Prepare 4 µM master mix of protein in buffer (20 mM Hepes pH 7.4, 150 mM NaCl, 5% glycerol) and add 1:1000 dilution of SYPR Orange.

    2.

    Aliquot 20 µL of the master mix into a white 96 well plate.

    3.

    Add DMSO or small molecule binder with a final concentration of 10 µM.

    4.

    Seal plate, mix well and centrifuge 30 sec at 500xg.

    5.

    Place plate into MX3005P real-time PCR instrument.

    6.

    Measure fluorescence with excitation and emission filters set to 465 and 590 nm while gradually increase temperature 3K/min during 71 cycles from 25 to 95°C.